Keresés

Menü

MSR 255 Multi-channel measuring system

Home » Product Range » Special (miniature) data loggers » MSR 255 Multi-channel measuring system

MSR 255 Multi-channel measuring system

Robust, multifunctional data acquisition device with LCD display

MSR_255
MSR_255



MSR255: Robust, multifunctional device with LCD display

With up to 5 different sensors, 4 additional analog inputs, and an LCD display, the MSR255 provides maximum flexibility and convenience.

5 sensors, 4 analog inputs

The configuration of the MSR255 is determined by the measurement tasks set by the user: it is possible to use temperature, humidity, air pressure, and light sensors (installed inside or outside the housing via cable) as well as 3-axis acceleration sensors. Accelerations can be measured up to 3200 Hz, with approximately 50 Hz frequency digital filtering. In addition, it can measure on 4 external analog voltage measurement channels or via sensors. The analog inputs have signal output, multi-output switchable power supply, and an input for starting/stopping measurements.

The memory capacity of the MSR255 is sufficient for 2 million measurable values. Thanks to the high-capacity lithium-polymer battery, the device is suitable for long-term data acquisition applications.

The 4-line backlit LED display can be configured into 4 unique positions. The saved data can be easily and quickly accessed via USB connection on a PC or laptop.

  1. General technical data of MSR255.
  2. Selectable measured parameters (external and internal).
  3. Do you need 4 additional inputs to connect sensors from a third party?

1. General technical data of MSR255.

HousingTechnical dataProduct code
Datenlogger MSR255: Robustes Multitalent mit LC-Anzeige, staubdicht IP 60

IP 60 standard
not sealed

Housing:Aluminum-coated industrial housing, IP60 standard, optional IP67 protection, DIN rail mountable TS 35
Size and weight:78 x 62 x 38 mm, approx. 222 g
 
Memory:More than 2 million measured values
 
Operation:Two-button function selection and data collection control
Display4-line LCD matrix
LED3-color LED indicator for data recording/alert/charging
 
Built-in sensor:Selectable temperature, relative humidity, air pressure, light, or 3-axis acceleration/position sensor
Measurement speed:1/s -1/12h (acceleration up to 3200/s)
Storage speed:1/s -1/12h (acceleration up to 50/s)
 
Power supply:  Lithium-polymer battery 2300 mAh | USB charger
Connector:USB (Mini-B)
PC software:Free installation, Reader, Viewer, and Online software (Windows XP/Vista/7/8/10) for data recording and evaluation.
 
Operating conditions:  -20 °C to +65 °C
Storage conditions:
  • +5…+45°C (ideal for battery storage)
  • 10…95% relative humidity, non-condensing environment
 
Certification:The MSR255 complies with the EU RoHS / WEEE standard.

255B10

Error

Data logger MSR255: Robust all-rounder with LC display, waterproof IP 67

IP 67 waterproof
sealed


255B11

Error

Selectable measured parameters (internal and external).

The MSR255 can be equipped with 5 sensors. The following external sensors can replace the internal ones: temperature, relative humidity, air pressure, and light.

Measured parameter /
Measurement/storage
Operating range /
Accuracy
Internal
sensor
External sensor (Based on cable length)
0.20 m1.00 m1.60 m
Temperature

1/s – 12 h

external: -55…+125 °C
internal: -20…+65 °C

±0.5 °C (-10…+65 °C)
±2 °C (-55…+125 °C)
T2
Error
FT2/020
Error
FT2/100
Error
FT2/160
Error
Relative humidity
(with built-in temperature sensor)

1/s – 12 h

0…100 % Rel. humidity
external: -20…+85 °C
internal: -20…+65 °C

±2 % Rel. humidity
(10…85 %, 0…+40 °C)
±4 % Rel. humidity
(85…95 %, 0…+40 °C)
H
Error
FH/020
Error
FH/100
Error
FH/160
Error
Air Pressure absolute,
(with built-in temperature
sensing)

1/s to every 12 h

0…2000 mbar absolute
external: -20…+85 °C
internal: -20…+65 °C

±2.5 mbar
(750…1100 mbar absolute,
+25 °C)
P
Error
FP/020
Error
FP/100
Error
FP/160
Error
0…14 bar absolute
-20…+65 °C

±50 mbar
(1…10 bar abs., +25 °C)
P2
Error
FP2/020
Error
FP2/100
Error
FP2/160
Error
3-axis Accelerometer

Acceleration: 50/s (3200/s)
Position:
1/s – 12 h

±15 g
-20…+65 °C

±0.15 g (0…5 g, +25 °C)
±0.25 g (5…10 g, +25 °C)
±0.45 g (10…15 g, +25 °C)
A
Error
FA/020
Error
FA/100
Error
FA/160
Error
Light

1/s - 12 h

0...65,000 lx
max. sensitivity
500 nm
L
Error
FL/020
Error
FL/100
Error
FL/160
Error

3. Do you need 4 additional inputs to connect sensors from a third party?

The MSR255 can be expanded with 4 analog inputs. These inputs can be used for a wide range of external sensors. If necessary, sensors from other manufacturers can be powered from the data logger's battery. The power supply can be automatically switched on before measurements are taken, helping to account for the sensors' start-up time.

Analog inputInputInput configurationProduct code
Screw terminal for 4 analog inputs
Includes alarm output
and input for starting and stopping data logging

Measurement/storage speed:
1/s to 1/12h

4 analog Input
(Resolution: 12 Bit)
0...3.0 V - Standard E3333
Error
4 analog Input
(Resolution: 12 Bit)
Select the number of inputs for the
appropriate configuration:

(1) 0...20 mA

(2) 4...20 mA

(3) 0…3.0 V

(4) 0.5…4.5 V / 0…5.0 V

(5) 0…10.0 V / 1.0…6.0 V

(6) 0…12.0 V

(7) 0…24.0 V

Exxxx
Error

Selected:

Device number:

No further information available.